6 Osmetech OPTI™ Repair Procedures
Service Manual – Osmetech OPTI™ Analyzer
6-9
Installation
To reinstall the assembly, follow the removal instructions in reverse order.
Before tightening the mounting screws, make sure to center the display over the
window.
Verification
To test the display go to
DIAG2 – Displ
and press
ENTER
. Check if all dots are
dark. If dots remain blank, the display assembly must be replaced.
Keypad
Removal
The keypad on the Osmetech OPTI 1 can be removed by first opening the lower
portion of the analyzer. The flex cable from the keypad will now need to be
disconnected from the display board. Use a knife to lift off the keypad from the
analyzer housing.
Installation
Before replacement with a new keypad, the surface must be cleaned. Use alcohol
to remove any adhesive which may be on the surface. The surface must be clean
and dry before installing a new keypad. Take the new replacement keypad and
remove the backing to expose the adhesive side of the keypad. Connect the
keypad cable to the display board. Feed the cable through the slot in the analyzer
housing. Carefully apply the new keypad to ensure proper alignment on the
analyzer housing.
Caution: The keypad is attached to the housing of the analyzer with an adhesive
backing and can not be re-used once removed from the analyzer.
Verification
To test the keypad go to
DIAG2 - Keypd
and press
ENTER
. Check if all the
buttons result in the correct response. The keypad must be replaced in case of any
malfunction or sluggish response.