2 General Description
2-4
Service Manual – Osmetech OPTI™ Analyzer
Input Values
Range *
Default
Value
Osmetech
OPTI 1
Osmetech
OPTI CCA
Rate (f)
0 to 155
0
3
Flow Rate (Liter Flow) (FR)
0 to 300
0
3
Inspiratory/Expiratory Ratio
(I/E Ratio)
1:9.9 0
3
Bi-Level Pressure
0 to 99.9 / 0 to 99.9
0 / 0
3
User Field
blank to 999999999
Off
3
User Field 2
blank to 999999999
Off
3
User Field 3
blank to 999999999
Off
3
Calculated Values
Range
Display
Resolution
Units
Osmetech
OPTI 1
Osmetech
OPTI CCA
Base excess in vitro
BE
-40 - +40
0.1
mmol/L
9
9
Base excess in vivo
BEecf
-40 - +40
0.1
mmol/L
9
9
Base excess actual
BEact
-40 - +40
0.1
mmol/L
9
Buffer base
BB
0 - 100
0.1
mmol/L
9
9
Actual bicarbonate
HCO
3
-
1 - 200
0.1
mmol/L
9
9
Total CO
2
TCO
2
1 - 200
0.1
mmol/L
9
9
Standard bicarbonate
stHCO
3
- 1 - 200
0.1
mmol/L
9
9
Standard pH
stpH
6.5 - 8.0
0.001
pH units
9
9
Oxygen saturation
O
2
sat
0 - 100
0.1
%
9
9
Oxygen content
O
2
ct
0 - 56
0.1
vol%
9
9
Hydrogen ion concentration
cH
+
1000 - 10
0.1
nmol/L
9
9
Alveolar-arterial oxygen
partial pressure difference
AaD O
2
0 - 800
0.1
mmHg
9
9
P50
P50
15 - 35
0.1
mmHg
9
Anion Gap
AnGp
3 - 30
0.1
mmol/L
9
Normalized Ca++
nCa++
0.1 - 3.0
0.1
mmol/L
9
Hematocrit, Calculated
Hct(c )
15 - 75
1
PCV%
9