5 Troubleshooting
Service Manual – Osmetech OPTI™ Analyzer
5-1
5 Troubleshooting
Osmetech OPTI™ System
The Osmetech OPTI™ 1 and Osmetech OPTI™ CCA analyzers are equipped
with sensors and diagnostic software that control, monitor and detect analyzer
system status. Operator messages indicating system status are displayed to
provide the operator with useful information to determine the cause of detected
system errors or malfunctions. System errors are also stored in memory and an
error report can be printed that lists the last 50 status messages.
Displayed Alarms
There are four types of displayed system status messages:
SYSTEM
STOP
ALARMS
These alarms indicate system conditions that must be resolved before system
operation can be continued.
SYSTEM ERROR ALARMS
These errors occur during sample analysis and are specific to the current sample
being analyzed. Error alarms indicate the status of the current measurement or
additional required operator entry.
SYSTEM WARNING MESSAGES
System warning messages notify the operator of conditions requiring operator
intervention to complete the current measurement.
SYSTEM INFORMATION MESSAGES
These messages describe operator instructions or confirm completion of operator
selections.
The following section describes each displayed message and the causes and
conditions that generate this message. Further in the chapter probable causes and
fault isolation are provided for each status message.