5 Troubleshooting
Service Manual – Osmetech OPTI™ Analyzer
5-25
Osmetech OPTI Troubleshooting - 4
QC Values
out of Range
pH Value
low
pH Value
high
pCO2 Value
high
pCO2 Value
low
Check
sampling
technique
(NO syringe!)
Replace
Cassettes
Replace
Gas Bottle
Install
current
SW
Version
Check
Sampling
Technique
(NO Syringe!)
Replace
Cassettes
Check Sampl.
Technique
(NO Syringe!)
pO2 Value
high
pO2 Value
low
Obtain Info:
- QC Lot #
- Assay Ranges
- QC Stats
- SW Version
Check for
Corr Factors
off
('-' on header)
Check Time/
Date setting
Check baro
Frequent
unstable pH?
Long cal.
(blinking '*')?
New cass.
fix problem?
Return Instr.
(Suspect Temp.
T1, T2 problem)
Replace
Cassettes
ok
NO
YES
YES
NO
NO
YES
Replace
Gas Bottle
Replace
Cassettes
Check
Technique
Check for voids
and bubbles
Controls not at
room temp?
(refrigerated?)
Controls not
at room
temp?
(hot?)
Check
Ampoule
handling
Establish
custom assay
ranges
Check
Sampl.
Technique
(NO Syringe!)
Check
storage temp
of Control
(18 - 27°C)
Replace
Cassettes
High Altitude?
YES
NO