2 General Description
2-28
Service Manual – Osmetech OPTI™ Analyzer
Diagnostic Menu – Osmetech OPTI CCA
Fig. 2-16: Diagnostic Menu
DIAG1:Baro Batt Temp Gas
Vers Reports BCode DIAG2
DIAG3: GValve Flow Disp
VDrive Pump keypad DIAG1
DIAG2: Printr Optics LEDs
RS232 IR Fan FSET DIAG3
DIAG1: - Barometer
Baro: 731.9 mmHg
DIAG1: - Battery
Battery Voltage: 12.9 V
DIAG1: - Temp. T3:31.48
T1:37.31 T2:37.01
DIAG1: - Gas
Gas Pressure: High
DIAG1: - Cleaning
Remove PumpCart<ENTER>
DIAG2: - Printer Test
DIAG2: - Keypad
Key:
DIAG2: - Interface
RS232-Jumper2-3 <ENTER>
DIAG3: - Valve
Valve: ON OFF
DIAG3: - VDrive
Position: 0
DIAG3: - Pump
Pump Speed: 800
RUN: SRC Controls
Data TimeDate Setup Diag
DIAG1: - Gas
Purge <ENTER> or <ESC>
DIAG1: - Cleaning
Place Syringe <ENTER>
DIAG1: - Cleaning
Perform Wash <ENTER>
DIAG1: - Cleaning
Reinsert Pump <ENTER>
Password <ENTER>
DIAG1: - Version
Version: ROPX2.24
DIAG1: - Version
Optics: OPTIC2H
DIAG1: - Report
Errors Cals SRCs Config
Error Report
Calibration Report
SRC Report
Configuration Report
BC:
DIAG2: - Test Optics
DIAG2: - Optics Test xxx
Test in Process ..
DIAG2: - Optics Test
Remove SRC
DIAG2: - Optics Test
Insert SRC <Enter>
DIAG2: - LEDs LT1:0 LT2:0
L1: 2614 L2: 2988 CD: 519
DIAG2: - IR
IR Loopback Test: PASS
DIAG2: - Fan
Cooling Fan: Off On
DIAG3: - Flow Test
Insert Cassette <ENTER>
DIAG3: - Flow Test
Test in Process ......
DIAG1: - Flow
Pass: <ENTER>
DIAG1: - Flow Test
Discard Cassette
DIAG2: - Display Test