![Osmetech Opti 1 Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/osmetech/opti-1/opti-1_service-manual_1655869032.webp)
2 General Description
Service Manual – Osmetech OPTI™ Analyzer
2-21
Diagnostic Menu – Osmetech OPTI 1
DIAG1:Baro Batt Temp Gas
Clean Flow TestOptics >
DIAG3: Valve VDrive Pump
LEDs Mem Vers Reports >
DIAG2: Printr Keypd Displ
Barcode RS232 Cover Fset >
DIAG1: - Barometer
Baro: 731.9 mmHg
DIAG1: - Battery
Battery Voltage: 12.9 V
DIAG1: - Temp. T3:31.48
T1:37.31 T2:37.01
DIAG1: - Gas
Gas Pressure: High
DIAG1: - Cleaning
Remove PumpCart<ENTER>
DIAG1: - Flow Test
Insert Cassette <ENTER>
DIAG1: - Test Optics
DIAG2: - Printer Test
DIAG2: - Keypad
Key:
DIAG2: - Display Test
DIAG2: - Barcode
Barcode:
DIAG2: - Interface
RS232-Jumper2-3 <ENTER>
DIAG2: - Cover
Cover: Closed
DIAG3: - Valve
Valve: ON OFF
DIAG3: - VDrive
Position: 0
DIAG3: - Pump
Pump Speed: 75
DIAG3: - LEDs LT1:0 LT2:0
L1: 2614 L2: 2988 CD: 519
DIAG3: - Memory
Memory OK
DIAG3: - Version
Version: AOPX1.10
DIAG3: - Report
Errors Calibns SRCs Config.
DIAG3: - Version
Optics: OPTIC2B
Osmetech OPTI 1
Error Report
RUN: SRC Controls
Data TimeDate Setup Diag
Osmetech OPTI 1
Calibration Report
Osmetech OPTI 1
SRC Report
DIAG1: - Gas
Purge <ENTER> or <ESC>
DIAG1: - Cleaning
Place Syringe <ENTER>
DIAG1: - Cleaning
Perform Wash <ENTER>
DIAG1: - Cleaning
Reinsert Pump <ENTER>
DIAG1: - Flow Test
Test in Process ......
DIAG1: - Flow 4.54 5.70
Pass: <ENTER>
DIAG1: - Flow Test
Discard Cassette
DIAG1: - Test Optics
Test in Process ..
DIAG1: - Test Optics
Remove Cassette
DIAG1: - Optics Test
Insert Cassette <Enter>
Password <ENTER>
Osmetech OPTI 1
Config. Report
Fig. 2-9: Diagnostic Menu