6 Osmetech OPTI™ Repair Procedures
Service Manual – Osmetech OPTI™ Analyzer
6-19
Analyzer Housing
The main analyzer housing is comprised of two (2) parts, the lower and upper
housing. The assemblies located inside the analyzer can be removed. To gain
access to these assemblies, the analyzer housings must be separated. To open the
analyzer housing, first remove the battery on the left side of the analyzer. Turn
the analyzer over to access the seven (7) screws securing the two housing
sections together. Remove the screws and separate the upper and lower housing
sections by flipping the lower housing open on the side of the carrying handle.
Caution: When turning the analyzer over to access the lower housing section,
place the analyzer on a padded surface to prevent scratches or other damage to
the upper housing section.
Figure 17: Opened Osmetech OPTI CCA Analyzer
Main Board
Assembly
Battery
Receptacle
SMC motor
Peristaltic
Pump Motor
SMC
Storage
Compartment
Gas Manifold
Module
Bar Code Reader
Cooling Fan