5 Troubleshooting
Service Manual – Osmetech OPTI™ Analyzer
5-11
SYSTEM INFORMATION MESSAGES
No SRC Record
(Level X)
SRC Database Deleted
(Level X)
Please Wait
Printing Stopped
Please Leave Cassette &
Close the Cover
Terminate Operation: N Y
When the operator has requested a printed report of the
SRC measurement statistics and no records are stored in
the SRC database this status message will be displayed.
The displayed message also identifies the applicable SRC
Level for which no SRC records exist.
When the Osmetech OPTI analyzer detects a new SRC
ID, the operator is prompted to delete all current SRC
data (password protected). As the operator enters the
correct password and replies
E
to delete the SRC data,
the analyzer prints the SRC statistical report and confirms
deletion of the SRC data with this displayed message.
While the Osmetech OPTI internal thermal printer is
printing a report, the operator may press
E
to abort
printing. The analyzer will display this message until
the current report is restarted or the operator presses
E
a second time to abort the print cycle.
Following a sample measurement, the cassette reservoir
is locked and the sample waste is sealed inside the
cassette. This reservoir locking prevents the
biohazardous waste from leaking and forms a sealed
container for the sample waste. If the SMC Cover is
opened prematurely in order to discard the cassette and
the valve drive has not been turned to the locking
position, this message is displayed. Close the cover and
wait for the analyzer to prompt for cassette removal.
During an operating sequence of the Osmetech OPTI,
the operator may select to abort the current operation.
To do this, press
E
and then “0”. The analyzer then
displays this message and waits for a key response (
J
or
E
) before terminating the current operation.