3 Test Programs and Diagnostics
Service Manual – Osmetech OPTI™ Analyzer
3-15
IR Test
The IR Loopback test is included to test the IR interface function. To perform
the test first select the IR test. Test results will be displayed indicating either
PASS or FAIL.
Fan Test
The FAN test allows the cooling fan to be turned ON and OFF using the analyzer
keypad. Turn the FAN ON and check for proper airflow at the fan vents on the
rear of the analyzer.
Note: The fan may only run for a short time. The software background task is
still active in this menu, turning the fan off if the internal temperature is low.
Fset
The Fset menu contains factory-set values stored in the EEPROM.