5 Troubleshooting
Service Manual – Osmetech OPTI™ Analyzer
5-19
Error Message
Possible Causes
Remedial Action
ERROR Possible Clot
Discard cassette <ESC>
LG 1 detected sample aspiration
and analyzer timed out before
sample reached LG 2.
Check sample cassette for clots.
Run properly heparinized
sample.
Check peri pump cartridge and
replace as necessary.
ERROR - Invalid Sample
Discard Cassette <ENTER>
During OPTI-trol measurement,
the sample detected was not a
blood sample.
Measure a blood sample after
OPTI-trol measurement.
WARNING Bad Sensor - pH
Continue <ENTER> or <ESC>
WARNING Bad Sensor - PCO2
WARNING Bad Sensor - PO2
WARNING Bad Sensor - Na
WARNING Bad Sensor - K
WARNING Bad Sensor - Ca
WARNING Bad Sensor - Cl
WARNING Bad Sensor - Glu
Air bubbles in cassette or void
over sensor. Possible poor
sample preparation or corrupt
optics giving < 7,000 counts.
Check for air bubbles in
cassette.
Remix sample and rerun with
new cassette.
Cycle analyzer power.
Open SMC cover and check
operation of Blue LED’s.
Replace the SMC Assembly.
WARNING - Unstable pH
Check Cassette
WARNING - Unstable PCO2
WARNING - Unstable PO2
WARNING - Unstable Na
WARNING - Unstable K
WARNING - Unstable Ca
WARNING - Unstable Cl
WARNING - Unstable Glu
Possible poor sample
preparation or air bubbles in
cassette.
Check for air bubbles in
cassette.
Remix sample and rerun with
new cassette.
WARNING – Gas Expires
Soon!
Press YES to Continue
Gas will expire in two weeks.
WARNING - No Pat. ID Entry
Continue <ENTER>
Patient ID is selected as a
required entry
Press <ENTER> to continue
and enter a Patient ID