5 Troubleshooting
5-8
Service Manual – Osmetech OPTI™ Analyzer
WARNING-Bad Sensor - pH
Continue<ENTER>or<ESC>
WARNING-Bad Sensor- CO2
Continue<ENTER>or<ESC>
WARNING-Bad Sensor - O2
Continue<ENTER>or<ESC>
WARNING-Bad Sensor- Na
Continue<ENTER>or<ESC>
WARNING-Bad Sensor - K
Continue<ENTER>or<ESC>
WARNING-Bad Sensor – Cl-
Continue<ENTER>or<ESC>
WARNING-Bad Sensor– Ca++
Continue<ENTER>or<ESC>
WARNING-Bad Sensor- Glu
Continue<ENTER>or<ESC>
These warning messages indicate a single sensor has an
output of less than 7,000 counts. The warning message may
appear either during the calibration or sample measurement
process. If this message occurs during sample measurement,
check sample preparation. These warning messages are
displayed and are specified for each measurement parameter.
When this warning is displayed no sample measurement
results will be reported for the parameter identified, however
results will be reported for the other parameters.