Manual Version e1-4.3.0.ae-R3.
Page
91
Figure 39. Maintenence menu.
The e1 Series will automatically test the pressure of the bottle. This provides a test method to qualify
XeF
2
pressure at any point in time as shown below.
Please note that the 2 Torr value used below is an illustration and your organization may decide to
choose another pressure value to make the bottle change determination.
For example, if the pressure displayed is above approximately 2 Torr, click NO and you will be returned
to the maintenance menu. If the pressure is below approximately 2 Torr, click YES, and the bottle
change will continue.
If you have chosen to change the XeF
2
bottle, you will be instructed to close the manual valve on the
bottle. Once you have closed the bottle valve, click OK as shown below.
Содержание SPTS Xactix e1 Series
Страница 1: ...Manual Version e1 4 3 0 ae R3 2015 04 Xactix e1 Series System Manual ...
Страница 38: ...Manual Version e1 4 3 0 ae R3 Page 36 Figure 23 Shroud ventillation connection Shroud ventilation connection ...
Страница 40: ...Manual Version e1 4 3 0 ae R3 Page 38 Figure 26 Showerhead stop installed ...
Страница 48: ...Manual Version e1 4 3 0 ae R3 Page 46 Figure 29 Chamber ventilation shroud moved to back ...
Страница 49: ...Manual Version e1 4 3 0 ae R3 Page 47 Figure 30 Chamber ventilation shroud pulled forward ...
Страница 83: ...Manual Version e1 4 3 0 ae R3 Page 81 ...
Страница 86: ...Manual Version e1 4 3 0 ae R3 Page 84 ...
Страница 101: ...Manual Version e1 4 3 0 ae R3 Page 99 Figure 46 Gax box heater ...
Страница 103: ...Manual Version e1 4 3 0 ae R3 Page 101 Figure 48 Vacuum interlock switches Chamber vacuum switch Main vacuum switch ...
Страница 109: ...Manual Version e1 4 3 0 ae R3 Page 107 9 Set number of purge cycles to 0 and Hard Pump Time to 10 minutes ...