Manual Version e1-4.3.0.ae-R3.
Page
103
Figure 50. Valve diagram.
WARNING
MANUAL OPERATION OF THE VALVES COULD LEAD TO
DAMAGE TO THE EQUIPMENT OR HARM TO PERSONNEL.
MANUAL VALVE OPERATION IS INDEPENDENT OF
INTERLOCKS AND SHOULD ONLY BE PERFORMED BY
TRAINED PERSONNEL.
In service or maintenance circumstances, only to be performed using trained personnel, where manual
operation of the valves is necessary, manual valve operation can be accomplished using a small allen
wrench or other small, pointed object. The valve manifold, see Figure 49, has 16 buttons that can be
pressed to force a valve to open. For reference, the valve diagram for the system is shown in Figure
50.
6.15 Ventilation shroud interlock
The ventilation shroud is interlocked via a safety switch/lock mechanism as shown in Figure 51. This
mechanism is controlled via the PC and the input/output signals related to this interlock are
communicated through the device shown in Figure 52.
If it is necessary to override the Ventilation shroud interlock, a pair of keys is provided that are used to
unlock the shroud.
Содержание SPTS Xactix e1 Series
Страница 1: ...Manual Version e1 4 3 0 ae R3 2015 04 Xactix e1 Series System Manual ...
Страница 38: ...Manual Version e1 4 3 0 ae R3 Page 36 Figure 23 Shroud ventillation connection Shroud ventilation connection ...
Страница 40: ...Manual Version e1 4 3 0 ae R3 Page 38 Figure 26 Showerhead stop installed ...
Страница 48: ...Manual Version e1 4 3 0 ae R3 Page 46 Figure 29 Chamber ventilation shroud moved to back ...
Страница 49: ...Manual Version e1 4 3 0 ae R3 Page 47 Figure 30 Chamber ventilation shroud pulled forward ...
Страница 83: ...Manual Version e1 4 3 0 ae R3 Page 81 ...
Страница 86: ...Manual Version e1 4 3 0 ae R3 Page 84 ...
Страница 101: ...Manual Version e1 4 3 0 ae R3 Page 99 Figure 46 Gax box heater ...
Страница 103: ...Manual Version e1 4 3 0 ae R3 Page 101 Figure 48 Vacuum interlock switches Chamber vacuum switch Main vacuum switch ...
Страница 109: ...Manual Version e1 4 3 0 ae R3 Page 107 9 Set number of purge cycles to 0 and Hard Pump Time to 10 minutes ...