Figure 35. Electrical plug lockout and tagout.
6.2.2
Gas Supply Disconnect and Lockout / Tagout
When working on the system the user should ascertain if there is any danger to themselves or the
system from the pressurized gasses delivered to the system. The required, customer supplied
regulators at the gas supply should be used to turn off the gas supply to the system and be the
customer supplied lockout / tagout procedure used to prevent accidental pressurization of the system.
The N
2
regulator inside of the gas box should also be reduced to zero psi (note that the N
2
regulator
inside of the system is not relieved so reducing the regulator will only prevent additional gas from
coming into the system – N
2
in the lines could be relieved by manually opening valve 8 after the
chamber has completed an unload operation – see Section 5.2.1). The CDA regulator on the rear of the
system should also be reduced to zero psi.
Содержание SPTS Xactix e1 Series
Страница 1: ...Manual Version e1 4 3 0 ae R3 2015 04 Xactix e1 Series System Manual ...
Страница 38: ...Manual Version e1 4 3 0 ae R3 Page 36 Figure 23 Shroud ventillation connection Shroud ventilation connection ...
Страница 40: ...Manual Version e1 4 3 0 ae R3 Page 38 Figure 26 Showerhead stop installed ...
Страница 48: ...Manual Version e1 4 3 0 ae R3 Page 46 Figure 29 Chamber ventilation shroud moved to back ...
Страница 49: ...Manual Version e1 4 3 0 ae R3 Page 47 Figure 30 Chamber ventilation shroud pulled forward ...
Страница 83: ...Manual Version e1 4 3 0 ae R3 Page 81 ...
Страница 86: ...Manual Version e1 4 3 0 ae R3 Page 84 ...
Страница 101: ...Manual Version e1 4 3 0 ae R3 Page 99 Figure 46 Gax box heater ...
Страница 103: ...Manual Version e1 4 3 0 ae R3 Page 101 Figure 48 Vacuum interlock switches Chamber vacuum switch Main vacuum switch ...
Страница 109: ...Manual Version e1 4 3 0 ae R3 Page 107 9 Set number of purge cycles to 0 and Hard Pump Time to 10 minutes ...