DMTA-10040-01EN, Rev. E, February 2018
Chapter 5
214
32. Depending on your requirements, press the DISP menu key (
) again, and
then select your preferred
D ERASE
(A key) or
PERSIST
(B key) values.
33. Press the ALARM menu key (
), and then set the
ALARM 1
(A key) to
MIX
POS
.
34. Press the ALARM menu key (
) again, and then set the
BOTTOM
(C key) to
50 %, the
LEFT
(D key) to
−
10 %, and the
RIGHT
(E key) to 110 %.
35. Press the FULL NEXT key (
) to toggle to the full-screen mode.
36. Place the probe on position 1, and then perform a final scan of both fastener rows
(see Figure 5-133 on page 214).
Figure 5
‑
133 The scan results for both fastener rows
To fine-tune the instrument settings
As desired, set the alarm box parameters, horn, or external horn (louder). For
more details about alarms, see “Alarm Menus” on page 292.
The list of all parameters is shown in Figure 5-134 on page 215.
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...