DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
203
Make sure that the guides are properly adjusted for both fastener rows. Failure to
properly center the probe will lead to a very difficult calibration and possibly an
unusable setup.
To set the initial NORTEC 600 configuration
1.
Connect the probe and the cable to the PROBE connector on the NORTEC 600.
2.
Press the ADV SETUP menu key (
) once, and select
APPL SELECT
(A key)
to open the application selection menu. Select
Dual Frequency Sub
‑
Surface
with
the knob, then press
to accept (see Figure 5-115 on page 203).
Figure 5
‑
115 The Dual Frequency Sub
‑
Surface application
3.
Press the MAIN FILTER menu key (
) four times, and then set the
LO PASS
(B key) to 50 Hz using the knob.
4.
Press the MAIN FILTER menu key (
) again, and then set the
PRB DRV
(A key) to
HIGH
.
5.
Press the DISP menu key (
) twice, and then set the
POSITION
(C key) to
BOT RGHT
.
IMPORTANT
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...