DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
183
and adjust the lift-off angle until it is oriented straight upward (90°) [see
Figure 5-84 on page 182].
Figure 5
‑
86 Adjusting the lift
‑
off angle to 90°
8.
Press the ERASE key (
), place the probe on the sample again (this time use a
larger lift-off or paint thickness value; for example, the thickness of three business
cards), and then press the GAIN key (
), and decrease the gain until the live
dot is within the screen area and is at about 90 % of the screen height (see
Figure 5-87 on page 183).
Figure 5
‑
87 Adjusting the gain
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...