DMTA-10040-01EN, Rev. E, February 2018
Chapter 5
254
Always use combined horizontal and vertical gain (
H/V GAIN
).
Figure 5
‑
186 Adjusting the gain
6.
Place the probe in a defect-free area of the calibration standard near the 10 % ID
flaw, and then press the NULL foot switch.
7.
Scan the 10 % ID and 20 % OD flaws.
8.
Press and hold the REF SAVE (
) key until you hear a beep, and then press the
ERASE key (
) or ERASE foot switch.
This provides a reference image in a complimentary color on the screen for
reference purposes (see Figure 5-187 on page 255).
IMPORTANT