DMTA-10040-01EN, Rev. E, February 2018
Chapter 5
164
8.
Press the FREEZE key (
) to unfreeze the acquisition, place the probe between
two notches on the standard, press the A-LIFT NULL key (
), and then scan the
standard.
The resulting signal should resemble the example shown in Figure 5-59 on
page 164. If necessary, freeze the signal again and fine-tune the vertical gain or
angle.
Figure 5
‑
59 The results after scanning the standard
To fine-tune the instrument settings
1.
Depending on your requirements, set the alarm parameters, horn, or external
horn (louder). For more details about alarms, see “Alarm Menus” on page 292.
The alarm parameters shown Figure 5-60 on page 165 in can be used as
NEGATIVE
Box alarm.
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...