DMTA-10040-01EN, Rev. E, February 2018
Chapter 3
68
2.
Repeatedly press the FULL NEXT key (
) until
APPLICATION WINDOW AT
STARTUP
is highlighted.
3.
Use the knob to disable the function (select
OFF
) or enable the function (select
ON
).
4.
Press
to return to the main inspection screen.
3.8
Enabling Knobless Entry for Harsh Environments
The knobless function allows the NORTEC 600 to be used in radioactive or harsh
environments where the instrument has to be put inside a bag, which makes it
difficult to use the knob.
If the knobless entry function is enabled, the frequency, angle, and the combined
horizontal and vertical gain settings can be increased or decreased using keypad
entries instead of the knob. The knobless entry function is enabled using the
SYSTEM
SETUP
menu. By default, the knobless entry function is set to
OFF
. For more
information on knobless entry, see “Knobless Entry” on page 78.
To enable the knobless entry function
1.
Press the ADV SETUP menu key (
) twice, then press the B key to enter the
SYSTEM SETUP
screen.
2.
Repeatedly press the FULL NEXT key (
) until
KNOBLESS
is highlighted.
3.
Use the knob to disable the function (select
OFF
) or enable the function (select
ON
).
4.
Press
to return to the previous screen.
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...