Chapter 5
168
2.
Ensure that the scanner motor is turned off before initial instrument nulling, place
the probe in the good hole, and then press the A-LIFT NULL key (
).
3.
Place the probe in the hole with EDM notches, and then adjust the index axis so
that the probe rotates freely next to a defect area.
Leave the scanner in the R position (rotation).
4.
Turn on the scanner motor, and then press the MAIN FILTER menu key (
)
twice, and set the
SCAN RPM
(E key) to
240
.
5.
Press the MAIN FILTER menu key (
), and then change the
FILT TYP
(C key)
to
FIG 6
(see Figure 5-64 on page 168).
This procedure uses the Figure 6 filter setting. Olympus recommends using the
Figure 6 filter setting unless an absolute probe is being used, in which case the
Figure 8 filter setting might provide better performance. For more information about
the Figure 6 versus Figure 8 parameters, see “Inspecting Fastener Holes with a
Rotating Scanner — NORTEC 600S and NORTEC 600D Models” on page 126.
Figure 5
‑
64 The Figure 6 filter
NOTE
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...