DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
247
Figure 5
‑
177 The tube scan result after moving the support ring (MIX)
4.
Place the probe in a defect-free area (see Figure 5-176 on page 246).
5.
Press the DISP menu key (
), followed by
CHANNEL
(B key), and then rotate
the knob until
FREQ 1
is displayed.
6.
Press the NULL foot switch, and scan the tube.
It should be easy to spot the flaw under the support ring (see Figure 5-178 on
page 247).
Figure 5
‑
178 The scan result with the support ring over the flaw in FREQ 1
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...