DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
241
Figure 5
‑
169 The frequency 2 scan signal of the 1.3 mm (0.052 in.) thru
‑
wall hole
with adjusted angle
8.
Press the FREEZE key (
) to enable signal acquisition.
9.
Place the probe in a defect-free area near the support ring, and then press the
NULL foot switch.
10. Scan the support ring to verify that the signal is below 80 % full-screen height
after the angle adjustment for the 1.3 mm (0.052 in.) thru-wall hole.
To record and cancel the support signal using AUTO MIX
1.
Press the DISP menu key (
), followed by
CHANNEL
(B key), and then rotate
the knob until the
MIX
menu is displayed.
2.
Press the MAIN FILTER menu key (
) three times.
3.
Place the probe in a defect-free area near the tube’s support ring, and then press
the NULL foot switch.
4.
Press
AUTO MIX
(B key), and slowly scan the support ring.
The
AUTO MIX
function is automatic and records the signal for 5 seconds after
AUTO MIX
(B key) has been pressed.
5.
Verify that
AUTO MIX
has acquired the support ring signal:
IMPORTANT
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...