To calibrate using MIX
1.
Press
CHANNEL
(B key), and then rotate the knob until
MIX
is displayed.
2.
Place the probe in a defect-free area of the calibration standard near the thru-wall
flaw, and then press the NULL foot switch.
3.
Scan the thru-wall flaw, and press the FREEZE key (
) when done.
4.
Press the MAIN FILTER menu key (
) three times.
The
MIX
menu appears (see Figure 5-203 on page 266).
Figure 5
‑
203 The MIX menu
5.
Press
MIX ANG
(mix angle, E key) and, if necessary, adjust the thru-wall flaw
signal until it is vertical.
6.
Press the FREEZE key (
) to enable signal acquisition.
7.
Press
AUTO MIX
(B key), and then scan the remainder of the tube.
The
AUTO MIX
function is automatic and records the signal for 5 seconds after
AUTO MIX
(B key) has been pressed.
8.
Verify that the
AUTO MIX
signal has been acquired:
IMPORTANT
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...