Repeat step 5 until you become familiar with the automatic lift-off function.
It may take some practice to obtain the proper timing with the automatic lift-off
function, but once mastered, this function will help you obtain a faster calibration.
7.
Press the ERASE key (
), then slowly scan the tapered area, and then, after
completing the scan, press the FREEZE key (
).
An example of the resulting scan is shown in Figure 5-93 on page 187.
Figure 5
‑
93 The scan of the tapered area
8.
Press the ANGLE key (
), and then adjust the lift-off angle so that it is