DMTA-10040-01EN, Rev. E, February 2018
Chapter 5
178
To sort metals by evaluating conductivity
1.
Connect the probe and the cable to the PROBE connector on the NORTEC 600.
2.
Press the ADV SETUP menu key (
) once, and then select
APPL SELECT
(A key) to open the application selection menu. Select
Surface Cracks
with the
knob, then press
to accept (see Figure 5-78 on page 178).
Figure 5
‑
78 The Surface Cracks application
3.
Press the DISP menu key (
), and set the
POSITION
(C key) to
BOT RGHT
.
4.
Press the MAIN FILTER menu key (
), and set the
FREQ
(A key) to 60 kHz.
5.
Set the
GAIN
(C key) to approximately 50 dB.
6.
Place the probe on the 45 % conductivity sample (see Figure 5-79 on page 179),
and then press and hold the A-LIFT NULL key (
) to activate the automatic
liftoff function.
After a brief moment, the NORTEC 600 beeps and displays
LIFT PROBE
at the
top of the screen. When this text is displayed, lift the probe into the air, and wait
for the message to disappear.
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...