DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
179
Figure 5
‑
79 The section of the standard used for conductivity evaluation
7.
Repeat step 6 until you become familiar with the automatic liftoff function.
It may take some practice to obtain the proper timing with the automatic lift-off
function, but once mastered, this function will help you obtain a faster calibration.
8.
Place the probe successively on the 17 %, 32 % and, 45 % conductivity samples,
and then press the FREEZE key (
) when done (see Figure 5-80 on page 179).
Figure 5
‑
80 The signals from different conductivity samples
9.
Press the ANGLE key (
), and then adjust the lower signal (45 %
conductivity) so that it is horizontal (see Figure 5-81 on page 180).