5.4 Command Descriptions
FRA5014
5-14
5.4.2.1 Common commands
*CLS
Description
Clears the status of the following.
Standard event status register
Operation event register
Overload event register
Status byte …
Note (see below)
Error queue
Parameters
None
Use example
*CLS
Note: The *CLS command does not directly clear the status byte register.
However, the status byte is indirectly cleared, except for the MAV bit and
RQS bit within that byte. The MAV bit can be indirectly cleared by
clearing the input buffer with a device clear command. The RQS bit can
be cleared by a serial polling operation to read the status.
*ESE <std_event_enable>
Description
Sets or queries contents of the standard event status enable register.
Parameter
<std_event_enable> Register contents, format <NR1>, range 0 to 255
For details, …
Use example
*ESE 255
Response
<std_event_enable>
Query example
*ESE?
Response example 255
*ESR?
Description
Queries contents of the standard event status register.
When queried, all bits of the standard event status register are cleared to 0.
Parameters
None
Response
Register contents, format <NR1>, range 0 to 255
For details,
Query example
*ESR?
Response example 128
*IDN?
Description
Queries model name, etc.
Parameters
None
Response
“Manufacturer name, model name, serial number, firmware version, format
<SRD>”
Query example
*IDN?
Response example “NF Corporation,FRA5014,9025257,Ver1.00”
Содержание FRA5014
Страница 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Страница 2: ......
Страница 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Страница 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Страница 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Страница 30: ......
Страница 48: ......
Страница 96: ......
Страница 107: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...