5.4 Command Descriptions
FRA5014
5-26
◼
Setting range for status-related 16-bit registers
The MSB (Most Significant Bit) in status-related 16-bit registers is pre-set as always “0”. When set
by a command, any value up to 65535 can be entered, but the MSB is always cleared to zero in the
target register. Accordingly, the response message for the query is always a value of 32767 or less.
STATus:OPERation:CONDition?
Description
Queries the operation condition register (OPCR).
Parameters
None
Response
Register contents, format <NR1>, range 0 to 32767
For details, …
Query example
STAT:OPER:COND?
Response example 0
STATus:OPERation:ENABle <opee>
STATus:OPERation:ENABle?
Description
Sets or queries the operation event enable register (OPEE).
Parameter
<opee> Register contents, format <NR1>, range 0 to 65535
For details, …
Setting example
STAT:OPER:ENAB 0
Response
<opee>
Query example
STAT:OPER:ENAB?
Response example 0
STATus:OPERation[:EVENt]?
Description
Queries the operation event register (OPER).
Parameters
None
Response
Register contents, format <NR1>, range 0 to 32767
For details, …
Query example
STAT:OPER?
Response example 0
Содержание FRA5014
Страница 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Страница 2: ......
Страница 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Страница 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Страница 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Страница 30: ......
Страница 48: ......
Страница 96: ......
Страница 107: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...