5.1 Preparations for Use
FRA5014
5-4
5.1.4 Selection of remote control interface
For the FRA5014, select GPIB, USBTMC or RS-232 as the remote control interface. Only one of
these can be used at a time. Use the DIP switches on the rear panel to select the interface and the
communication settings. The factory setting for the DIP switches is that only SW7 is set to 1, and
all other switches are set to 0 (GPIB address 2).
Figure 5-1. DIP Switches on Rear Panel
The “SW2 = 1, SW3 = 0” setting (LAN) is reserved internally. Do not use this setting.
(Example 1)
•
Interface
GPIB
•
GPIB address
2
Set SW1 to SW6 and SW8 to “0” and SW7 to “1”.
1
0
SW
1
2
3
4
5
6
7
8
Содержание FRA5014
Страница 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Страница 2: ......
Страница 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Страница 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Страница 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Страница 30: ......
Страница 48: ......
Страница 96: ......
Страница 107: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...