4.1 Accept/Reject Criteria
FRA5014
4-2
4.1 Accept/Reject Criteria
During spot measurements, auto judgments are made as to whether or not the measured gain (dB)
and/or phase value is within the designated range, and the results can be output.
4.1.1 Accept/Reject range settings
To perform accept/reject judgments, set upper and lower limit values as accept/reject criteria for
each channel.
When lower limit
upper limit, no judgment is performed.
◼
Accept/Reject Gain upper limit Gain lower limit
Range of values
−
199.99 to 199.99 dB, resolution is 0.01 dB
Gain judgment upper limit
Set this using the remote command “SENS:DATA:SPOT:LIM:GAIN:MAX parameter 1,
parameter 2, parameter 3”.
Parameter 1: Sets gain judgment upper limit for CH2 & CH1.
Parameter 2: Sets gain judgment upper limit for CH3 & CH1.
Parameter 3: Sets gain judgment upper limit for CH4 & CH1.
Gain judgment lower limit
Set this using the remote command “SENS:DATA:SPOT:LIM:GAIN:MIN parameter 1,
parameter 2, parameter 3”.
Parameter 1: Sets gain judgment lower limit for CH2 & CH1.
Parameter 2: Sets gain judgment lower limit for CH3 & CH1.
Parameter 3: Sets gain judgment lower limit for CH4 & CH1.
◼
Accept/Reject Phase upper limit Phase lower limit
Range of values
−
180.00 to 180.00 deg, resolution is 0.01 deg
Phase judgment upper limit
Set this using the remote command “SENS:DATA:SPOT:LIM:PHAS:MAX parameter 1,
parameter 2, parameter 3”.
Parameter 1: Sets phase judgment upper limit for CH2 & CH1.
Parameter 2: Sets phase judgment upper limit for CH3 & CH1.
Parameter 3: Sets phase judgment upper limit for CH4 & CH1.
Gain judgment lower limit
Set this using the remote command remote command “SENS:DATA:SPOT:LIM:PHAS:MIN
parameter 1, parameter 2, parameter 3”.
Parameter 1: Sets phase judgment lower limit for CH2 & CH1.
Parameter 2: Sets phase judgment lower limit for CH3 & CH1.
Parameter 3: Sets phase judgment lower limit for CH4 & CH1.
Содержание FRA5014
Страница 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Страница 2: ......
Страница 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Страница 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Страница 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Страница 30: ......
Страница 48: ......
Страница 96: ......
Страница 107: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...