2.2 Installation
FRA5014
2-4
2.2.3 Rack mount
When the (optional) rack mount adapter is attached to the
FRA5014
, it can be inserted into a
19-inch IEC rack, EIA-standard rack, or JIS-standard rack.
Attach the rack mount adapter as described below in “
Figure 2-3 Attachment of Rack Mount
Figure 2-4 Attachment of Rack Mount Adapter (Inch Based)
”,
then insert it into the rack.
Note the following precautions when inserting the
FRA5014
into a rack mount.
•
Always install supports such as rack rails for the
FRA5014
.
Otherwise, the
FRA5014
may fall from the rack, causing possible injury to others or
damage to the
FRA5014
.
•
Make sure the rack has spaces to allow ample air circulation, or install a cooling fan.
The
FRA5014
’s air intake holes are on its bottom panel. If other units are mounted
above and below the
FRA5014
, leave a gap of at least
20 mm
under the
FRA5014
.
Overheating beyond the rated maximum temperature can shorten performance life and
otherwise damage the
FRA5014
.
If any feet have been attached to the rails, remove them as described below in “
”.
!
CAUTION