5.4 Command Descriptions
FRA5014
5-15
*OPC
Description
Specifies that “1” will be set to the OPC bit (BIT0) in the standard event status
register when execution of all overlap commands is complete.
Overlap command …
Description of *WAI command
Parameters
None
Setting example
*OPC
Response
Format <NR1>
“1” is returned when all overlap commands have been completed. However,
executing *OPC? does not clear the OPC bit in the standard event status register.
Use a device clear, *CLS, or *RST command to clear this register.
Query example
*OPC?
(Use example: overlap command ; *OPC? <PMT>
In this case, PMT: Program Message Terminator)
Response example 1
*RCL <setting_memory>
Description
Switches to (recalls) the data stored in setting memory for the settings to be
measured.
Parameter
<setting_memory> memory number, format <NR1>, range 0
Setting example
*RCL 0
*RST
Description
Resets the device, turns output off, and returns measurement-related parameters
to their initial values (full initialization). For details, …
Parameters
None
Setting example
*RST
*SAV <setting_memory>
Description
Copies (saves) the currently used settings to the specified setting memory.
Parameter
<setting_memory> Memory number, format <NR1>, range 0
Setting example
*SAV 0
*SRE <srq_enable>
Description
Sets service request enable register.
Parameter
<srq_enable> Register contents, format <NR1>, range 0 to 255
For details,
5.5.1 Overview of status system
Use example
*SRE 128
Response
<srq_enable>
Query example
SRE?
Response example 128
Содержание FRA5014
Страница 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Страница 2: ......
Страница 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Страница 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Страница 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Страница 30: ......
Страница 48: ......
Страница 96: ......
Страница 107: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...