7.1 Introduction
FRA5014
7-2
7.1 Introduction
The following maintenance tasks should be performed to maintain this equipment in opt imum
condition.
•
Operation check
Check whether equipment is operating correctly.
•
Performance test
Check whether equipment meets specified rated values.
•
Adjustments and calibration
If rated values are not met, ask NF Corporation to make
adjustments and restore performance.
•
Repair
If not improved, ask NF Corporation to identify problems and
make adjustments and/or repairs.
This manual describes performance test methods that are easy to carry out.
For more advanced tests, adjustments, calibration, or repair, contact NF Corporation or a sales
representative.
The following measuring instruments are needed for the performance test.
•
Oscilloscope
Frequency band: 10 MHz or higher
•
Frequency counter
Accuracy:
5
10
−
6
or higher
•
AC voltmeter
Accuracy:
0.2 %, actual RMS value, frequency band: 100 kHz
or higher
•
DC voltmeter
Accuracy:
0.1 % or higher
•
Resistance meter (tester)
Measurement range: 10 M
Ω
or higher
7.2 Routine Checks
Use the
FRA5014
where the installation satisfies the conditions for use.
For description of installation conditions,
“
Installation conditions
” in “
” above.
If the panel or case becomes dirty, wipe it clean with a soft cloth. For stubborn dirt, dip the cloth in
a neutral detergent and wring it well before wiping with it.
Avoid using organic solvents such as thinner or benzene, or chemically treated cloths, since these can
cause discoloration or clouding of product surfaces, or may remove the product’s outer finish .
7.3 Storage, Repackaging, and Shipping
Always store the
FRA5014
in a location that meets the installation conditions.
For installation conditions,
When repackaging for shipment, etc., use a strong box with plenty of room for padding, using
packing material that can withstand the equipment’s weight, and make sure the equipment is fully
protected.
During shipment, handle with care and avoid strong impacts.
Содержание FRA5014
Страница 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Страница 2: ......
Страница 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Страница 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Страница 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Страница 30: ......
Страница 48: ......
Страница 96: ......
Страница 107: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...