5.6 Cautions for Programming
FRA5014
5-36
■ Related commands or queries
STATus:OVERload:ENABle / STATus:OVERload:ENABle?
This sets or queries the overload event enable register.
Set “0” to clear the enable register to zero.
There are no other commands to clear these bits.
The register is cleared to 0 immediately after power-on.
STATus:OVERload [:EVENt]?
This queries the overload event register.
Even if the cause of the overload event is removed, the register will not be cleared .
When queried, the event register is cleared to zero.
The event register can also be cleared by the *CLS command.
The register is cleared to 0 immediately after power-on.
The parameters in setting messages or response messages for each register are values obtained by
adding the weights of all “1” bits.
5.6 Cautions for Programming
◼
Caution when transmitting commands
When transmitting a command, add a line feed (0AH) as a program message terminator to the end of
the transmit string. If LF is not added, the system may not correctly operate. Transmission of a
program message terminator or END message has not been specified separately from the command
itself, they will not be output. Although the line feed may be described as a new line, the binary code
is the same.
Содержание FRA5014
Страница 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Страница 2: ......
Страница 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Страница 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Страница 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Страница 30: ......
Страница 48: ......
Страница 96: ......
Страница 107: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...