QG
set gear constant
IMM
PGM
MIP
USAGE
♦
♦
Units
xx
-
none
SYNTAX
xx
QG
nn or xx
QG
?
PARAMETERS
Description
xx
[ int ] -
axis number
nn
[float] -
gear constant
Range
xx
-
1
to
MAX AXES
nn
-
0
to
2e9
or
?
to read present setting
nn
-
evolution / unit of measure
Defaults
xx
missing:
error 37, AXIS NUMBER MISSING
out of range:
error 9, AXIS NUMBER OUT OF RANGE
nn
missing:
error 38, COMMAND PARAMETER MISSING
out of range:
error xx2, PARAMETER OUT OF RANGE
DESCRIPTION
This command is used to set the gear constant for a Newport Unidrive
compatible programmable driver for DC servo axis. This command should be
used in conjunction with
QT
(tachometer gain) command.
The gear constant is defined as the number of revolutions the motor has to make
for the motion device to move one displacement.
This command must to be followed by the
QD
update driver command to take
affect.
RETURNS
If the “?” sign takes the place of
nn
value, this command reports the current
setting
REL. COMMANDS
SN
-
set displacement units
QD
-
update driver
QS
-
set microstep factor
QI
-
set motor maximum current
QV -
et average motor voltage
EXAMPLE
2QG?
|
read gear constant setting of axis # 2
0.3937
|
controller returns a
value of 0.3937 rev / unit for axis #2
2QG 0.25
|
set gear constant to 0.25 rev / unit for axis #2
2QT
7.0 |
set tachometer gain to 7 V/Krpm for axis #2
2QD
|
update programmable driver with latest settings for axis #2
3-114
Section
3
–
Remote
Mode
Содержание ESP300 Series
Страница 1: ...ESP300 Motion Controller Driver User s Manual...
Страница 4: ...ESP300 1999 EU Declaration of Conformity iv Preface...
Страница 29: ...This page is intentionally left blank Section 2 Modes of Operation 1 15...
Страница 30: ......
Страница 230: ...4 16 Section 4 Advanced Capabilities...
Страница 274: ...This page is intentionally left blank 6 6 Section 6 Servo Tuning...
Страница 290: ...This page is intentionally left blank B 6 Appendix B Trouble Shooting and Maintenance...