Chapter 7
Counters
©
National Instruments
7-29
X Series User Manual
Fig
u
re 7-26 shows an example of a sample clocked b
u
ffered two-signal
separation meas
u
rement.
Figure 7-26.
Sample Clocked Buffered Two-Signal Separation Measurement
Hardware-Timed Single Point Two-Signal
Separation Measurement
A hardware-timed single point (HWTSP) two-signal separation
meas
u
rement has the same behavior as a sample clocked b
u
ffered
two-signal separation meas
u
rement. Refer to the
Sample Clocked Buffered
Two-Signal Separation Measurement
section for more information.
Note
If an active edge on the Gate and an active edge on the AUX does not occ
u
r between
sample clocks, an overr
u
n error will occ
u
r.
Note
(NI USB-634
x
/635
x
/636
x
Devices)
X Series USB devices do not s
u
pport
hardware-timed single point (HWTSP) operations.
For information abo
u
t connecting co
u
nter signals, refer to the
Default
Counter/Timer Pinouts
section.
S
OURCE
Co
u
nter V
a
l
u
e
B
u
ffer
AUX
GATE
1 2
3
1
2
3
1
2
3
3
3
3
Sa
mple
Clock
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com