Chapter 11
Triggering
©
National Instruments
11-7
X Series User Manual
Figure 11-7.
Analog Edge Triggering with Hysteresis Falling Slope Example
•
Analog Window Triggering
—An analog window trigger occ
u
rs
when an analog signal either passes into (enters) or passes o
u
t of
(leaves) a window defined by two voltage levels. Specify the levels by
setting the window Top val
u
e and the window Bottom val
u
e.
Fig
u
re 11-8 demonstrates a trigger that asserts when the signal enters
the window.
Figure 11-8.
Analog Window Triggering Mode (Entering Window)
An
a
log Comp
a
ri
s
on Event
Hy
s
tere
s
i
s
High thre
s
hold
(Level + Hy
s
tere
s
i
s
)
Low thre
s
hold
(Level)
Fir
s
t
s
ign
a
l m
us
t go
ab
ove high thre
s
hold.
Then
s
ign
a
l m
us
t go
b
elow low thre
s
hold
b
efore An
a
log Comp
a
ri
s
on Event
ass
ert
s
.
Top
An
a
log Comp
a
ri
s
on Event
Bottom