Chapter 5
Analog Output
©
National Instruments
5-9
X Series User Manual
onboard clock, the generation res
u
mes as soon as the pa
u
se trigger is
deasserted, as shown in Fig
u
re 5-5.
Figure 5-5.
AO Pause Trigger with the Onboard Clock Source
If yo
u
are
u
sing any signal other than the onboard clock as the so
u
rce of
yo
u
r sample clock, the generation res
u
mes as soon as the pa
u
se trigger is
deasserted and another edge of the sample clock is received, as shown in
Fig
u
re 5-6.
Figure 5-6.
AO PauseTrigger with Other Signal Source
Using a Digital Source
To
u
se AO Pa
u
se Trigger, specify a so
u
rce and a polarity. The so
u
rce can
be one of the following signals:
•
PFI <0..15>
•
RTSI <0..7>
•
PXI_STAR
•
PXIe-DSTAR<A,B>
•
Co
u
nter
n
Internal O
u
tp
u
t
•
Co
u
nter
n
Gate
•
AI Pa
u
se Trigger (ai/Pa
u
seTrigger)
P
aus
e Trigger
Sa
mple Clock
P
aus
e Trigger
Sa
mple Clock
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com