![National Instruments DAQ X Series Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/national-instruments/daq-x-series/daq-x-series_user-manual_3551122071.webp)
Chapter 4
Analog Input
©
National Instruments
4-25
X Series User Manual
Figure 4-12.
Analog Input Timing Options
MIO X Series devices
u
se AI Sample Clock (ai/SampleClock) and
AI Convert Clock (ai/ConvertClock) to perform interval sampling. As
Fig
u
re 4-13 shows, AI Sample Clock controls the sample period, which is
determined by the following eq
u
ation:
1/Sample Period = Sample Rate
Figure 4-13.
MIO X Series Interval Sampling
AI Convert Clock controls the Convert Period, which is determined by the
following eq
u
ation:
1/Convert Period = Convert Rate
PFI, RT
S
I
PXI_
S
TAR
An
a
log Comp
a
ri
s
on
Event
20 MHz Time
bas
e
100 kHz Time
bas
e
PXI_CLK10
Progr
a
mm
ab
le
Clock
Divider
Progr
a
mm
ab
le
Clock
Divider
AI
Sa
mple Clock
Time
bas
e
AI Convert Clock
Time
bas
e
PFI, RT
S
I
PXI_
S
TAR
An
a
log Comp
a
ri
s
on Event
Ctr
n
Intern
a
l O
u
tp
u
t
S
W P
u
l
s
e
PFI, RT
S
I
PXI_
S
TAR
An
a
log Comp
a
ri
s
on Event
Ctr
n
Intern
a
l O
u
tp
u
t
AI Convert Clock
AI
Sa
mple Clock
100 MHz Time
bas
e
Ch
a
nnel 0
Ch
a
nnel 1
Convert Period
Sa
mple Period
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com