3 Operating Procedures
3.5.11 Pipette Tips Layout
Note:
The Assay Plate setting takes precedence over all other fluidics module settings. If
you change the assay plate after specifying Pipette Tips Layout, your tip settings will be
canceled and the layout will revert to default.
Molecular Devices recommends that you always use a full rack of tips (the default
setting). If you use less than a full rack of tips, it is imperative that you configure this
setting to correctly match the number and positions of the actual tips inserted in the tip
rack drawer, because the instrument cannot determine how many tips are actually
present. While it is possible to use a partial rack of tips, and to choose which tips are
available in the Pipette Tips Layout setting, do so with caution. Be sure that the layout
described in the software matches the actual tips inserted in the tip drawer.
Caution:
Molecular Devices recommends using a full rack of tips each time you perform
a transfer of fluid in Flex mode. If you mistakenly enable a pipetting function from a tip
that is not present, or if you enable more or fewer tips than are actually available, the
instrument can malfunction, potentially causing serious damage.
Pipette Tips Layout Settings
The dimension of this layout will be determined by the Assay Plate Type you select.
Highlight the tips you want use with the cursor. Tips must be contiguous, but do not need
to start in the ‘1’ column. You cannot choose a partial row or column, or a single well.
Important!
Selected tips must be contiguous.
68
FlexStation II Operator’s Manual – Rev. D
Содержание FlexStation II
Страница 6: ...FlexStation II Operator s Manual Rev D vi ...
Страница 92: ...4 Applications 86 FlexStation II Operator s Manual Rev D ...
Страница 102: ...5 Service and Maintenance Procedures 96 FlexStation II Operator s Manual Rev D ...
Страница 124: ...6 Troubleshooting Procedures 118 FlexStation II Operator s Manual Rev D ...
Страница 134: ...7 Appendices 128 FlexStation II Operator s Manual Rev D ...