6 Troubleshooting Procedures
6.4.2 Opening the Instrument
Caution:
Use caution during this procedure. Follow these instructions and any instruction
labels on the instrument exactly.
1 Power
off.
Turn the power off to the instrument. Disconnect the power cords from the instrument
and the cables connected to the computer.
2 Lift up instrument cover.
Press in the handle at the bottom of the top cover and lift the cover back off the
detection module, revealing the fluidics module.
Inside Front Panel
Fluidics Module
Detection Module
Inside Top Panel
Tip Rack Drawer
Location of Inside Panels and Tip Rack Drawer
3 Remove front panel of fluidics module.
DO NOT REMOVE COVER
UNTIL POWER IS DISCONNECTED.
FOR TRAINED SERVICE PERSONNEL ONLY.
Do not operate instrument
unless all covers are in place.
Turn the four quarter-turn screws and lift the front panel out.
You many need to inspect the interior of the module to determine if the pipettor was in
the process of a fluid transfer when the power was interrupted. The pipettor head should
be at rest in the uppermost (“home”) position. If not, there may be fluid in the tips that
needs to be removed. Instructions for these procedures follow.
104
FlexStation II Operator’s Manual – Rev. D
Содержание FlexStation II
Страница 6: ...FlexStation II Operator s Manual Rev D vi ...
Страница 92: ...4 Applications 86 FlexStation II Operator s Manual Rev D ...
Страница 102: ...5 Service and Maintenance Procedures 96 FlexStation II Operator s Manual Rev D ...
Страница 124: ...6 Troubleshooting Procedures 118 FlexStation II Operator s Manual Rev D ...
Страница 134: ...7 Appendices 128 FlexStation II Operator s Manual Rev D ...