5 Service and Maintenance Procedures
5.7 Replacing the Flash Lamp
The flash lamp should be replaced when an error message appears in SOFTmax PRO to
indicate replacement is necessary.
Caution:
This procedure requires trained service personnel. The instrument warranty
may be voided unless lamp is replaced according to these Instructions.
Caution:
Use caution during this procedure. Follow these instructions and any instruction
labels on the instrument exactly.
LAMP COVER
Disconnect power cord
before removing
Requires trained service personnel.
WARRANTY MAY BE VOIDED
unless lamp is replaced according to
instructions in the operators manual
under "Flash Lamp Replacement".
Lamp Cover
Flash Lamp Location
Tools Required
•
Allen and/or ball-driver hex wrenches 3/32", 9/64", 5/32"
•
#2 Phillips screwdriver
1 Turn off power.
Turn off the instrument power switch and unplug the power cord and the computer
cable from the rear of the instrument.
2 Remove lamp cover.
a
Remove the Phillips head screw on the top of the cover and lift the cover off
vertically.
b
Set the cover aside, and keep track of the screw.
Note:
There is an interlock switch, which is depressed by the lamp cover, which
interrupts power to the flash lamp automatically when the cover is removed.
92
FlexStation II Operator’s Manual – Rev. D
Содержание FlexStation II
Страница 6: ...FlexStation II Operator s Manual Rev D vi ...
Страница 92: ...4 Applications 86 FlexStation II Operator s Manual Rev D ...
Страница 102: ...5 Service and Maintenance Procedures 96 FlexStation II Operator s Manual Rev D ...
Страница 124: ...6 Troubleshooting Procedures 118 FlexStation II Operator s Manual Rev D ...
Страница 134: ...7 Appendices 128 FlexStation II Operator s Manual Rev D ...