3.10 Optimizing Fluorescence Assays in Spectrum Mode
The alternative cutoff wavelength choices are listed in the following table. The cutoff
value should be near the maximum emission wavelength (preferably between the
excitation wavelength and the maximal emission wavelength).
Automatic Cutoff Selection
Endpoint and Kinetic Modes
# Lambda
λ
(nm)
Emission
λ
(nm)
1 None
<
415
2 420
415-434
3 435
435-454
4 455
455-474
5 475
475-494
6 495
495-514
7 515
515-529
8 530
530-549
9 550
550-569
10 570
570-589
11 590
590-609
12 610
610-629
13 630
630-664
14 665
665-694
15 695
695-850
Emission Cutoff Filter Default Settings
Note:
For Spectrum mode, the default is Manual (that is, no automatic cutoff).
FlexStation II Operator’s Manual – Rev. D
83