6 Troubleshooting Procedures
3 Check
Preferences.
Go to Preferences in the Edit drop down menu of the software. View the Preferences
dialog box.
Edit Preferences Dialog Box
There are two settings in this dialog box that must be correct for the instrument to
operate.
a
Make sure the Serial Port selected is the same as the one on the computer to which
the instrument is physically connected. It may be COM1 or COM2.
b
Make sure the Serial Comm Speed is correctly selected. The communication speed is
9600 during the instrument’s initial warm up sequence, while it is first
communicating with the software.
Note:
The Serial Comm Speed automatically changes to 57,600 when you select Flex
mode from the Settings window.
If you follow these steps and are still having trouble starting up the system, contact
Technical Support.