1.6 Theory of Operation
Operation Modes
Endpoint
Kinetic
Spectrum
Well Scan
Flex
Operation
Type
Detection Detection Detection Detection Fl
Detection
Read Types
Fluorescence
Luminescence
TRF
Fluorescence
Luminescence
TRF
Fluorescence
Luminescence
TRF
Fluorescence
Luminescence
TRF
Fluorescence
only
System
Settings
Top read
Yes or No
Yes or No
Yes or No
Yes or No
No
Bottom read
Yes or No
Yes or No
Yes or No
Yes or No
Yes (always on)
Wavelength
Yes Yes Yes Yes Yes
Automix
before Yes Yes Yes Yes Yes
Automix between
No
Yes
No
No
Yes
Timing No
Yes
No
No
Yes
Wells
to
read
Yes Yes Yes Yes Yes
AutoCalibrate
Yes Yes Yes Yes Yes
Compound
source
No No No No Yes
Compound
transfer
No No No No Yes
Triturate
No No No No Yes
Pipette
tips
layout
No No No No Yes
Compound & tips
columns
No No No No Yes
AutoRead
Yes Yes Yes Yes Yes
Well Scan Editor
No
No
No
Yes
No
PMT
sensitivity Yes Yes Yes Yes Yes
Assay
plate
type Yes Yes Yes Yes Yes
Operation Modes and Features
Note:
The system performs limited luminescence and TRF readings.
Endpoint Mode
In Endpoint mode, as well as in Kinetic and Flex modes, you may select from one to
four excitation/emission pairs to obtain Relative Fluorescence Units (RFU) readings for
each well of a microplate.
In Endpoint mode, the fluidics module is not enabled.
For more information on this mode, please review the appropriate section in the
SOFTmax PRO User’s Manual.
Kinetic Mode
Kinetic analysis can be performed for a total run time of up to 99 hours. The kinetic
read interval depends upon the instrument setup parameters chosen in SOFTmax PRO,
but is limited to 2 hours and 45 minutes (165 minutes). At the end of a reading, rates
are reported as RFU/second for each well. Kinetic analysis has many advantages when
FlexStation II Operator’s Manual – Rev. D
25
Содержание FlexStation II
Страница 6: ...FlexStation II Operator s Manual Rev D vi ...
Страница 92: ...4 Applications 86 FlexStation II Operator s Manual Rev D ...
Страница 102: ...5 Service and Maintenance Procedures 96 FlexStation II Operator s Manual Rev D ...
Страница 124: ...6 Troubleshooting Procedures 118 FlexStation II Operator s Manual Rev D ...
Страница 134: ...7 Appendices 128 FlexStation II Operator s Manual Rev D ...