User
Guide
for
Axoclamp
900A
DC
Restore
When
selected
in
TEVC
or
dSEVC
modes,
DC
restore
increases
DC
Voltage
‐
Clamp
Gain
to
~1,000,000.
This
means
the
holding
level
is
restored
to
>
99%
of
the
programmed
value
in
<
100
ms.
Buzz
Capacitance
neutralization
increased
by
4
pF
for
a
programmable
time
(0.1
ms
to
500
ms).
This
is
also
available
remotely
via
a
tethered
buzz
box
(>
2.5
V
=
buzz).
Clear
(+/–)
Maximum
positive
or
negative
current—manual
up
to
0.5
s.
Step
Activate
Available
on
each
channel,
STEP
ACTIVATE
1
and
STEP
ACTIVATE
2
inputs
work
independently,
setting
external
timing
for
going
between
the
holding
level
and
a
programmable
pulse
with
external
timing
up
to
50
kHz.
External
timing
is
set
with
a
TTL
high
pulse
input
to
the
STEP
ACTIVATE
BNC
for
the
duration
of
the
step.
The
amplitude
of
the
step
is
set
by
the
Tuning
amplitude
in
current
clamp
and
by
the
Seal
Test
amplitude
in
voltage
clamp.
Blank
Activate
Available
on
rear
panel
BLANK
ACTIVATE
1
BNC
or
REMOTE
connector.
Used
for
blanking
response
to
external
stimuli
on
Channel
1
only.
170