Commander
Reference
The
Reset
To
Program
Defaults
command
sets
the
SoftPanel
to
its
default
values,
as
well
as
all
other
program
settings.
Detect
Oscillation
The
art
of
pipette
capacitance
neutralization
is
to
increase
the
neutralization
as
far
as
possible
without
provoking
oscillations
that
may
be
harmful
to
your
cell.
But
no
matter
how
carefully
you
compensate
capacitance
to
begin
your
experiment,
it
is
still
possible
to
experience
oscillations
later
in
an
experiment
because
electrode
properties
(
e.g.,
resistance,
junction
potential)
may
change
over
time.
Current
Clamp
The
Axoclamp
900A
provides
you
with
options
to
protect
your
cell
from
harmful
oscillations
during
a
current
‐
clamp
experiment
by
automatically
disabling
or
reducing
Pipette
Capacitance
Neutralization.
Use
the
Detect
Oscillation
functionality
as
follows:
Check
the
Detect
Oscillation
checkbox,
and
select
the
Disable
Neutralization
option.
Now,
increase
the
Pipette
Capacitance
Neutralization
until
you
reach
a
value
that
evokes
full
‐
scale
oscillations.
The
automatic
protection
circuit
will
work
quickly
to
disable
the
Pipette
Capacitance
Neutralization,
and
several
things
will
happen:
ο
To
the
right
of
the
“Disable…”
field,
a
small
icon
appears
briefly
to
display
repeated
images
of
a
sine
wave
that
is
reduced
to
a
flat
line.
ο
The
Pipette
Capacitance
Neutralization
feature
is
disabled
(box
is
unchecked).
ο
You
hear
an
audible
tone.
123