User
Guide
for
Axoclamp
900A
Bridge
Balance
in
the
Bath
Some
investigators
like
to
set
Bridge
Balance
in
the
bath,
before
attempting
to
impale
cells.
This
is
to
make
it
easier
to
see
when
a
cell
has
been
penetrated.
Check
the
Tuning
checkbox
and
set
the
parameters
to
–1
nA
and
50
Hz.
Observe
the
Membrane
Potential
on
Scaled
Output.
Press
the
Auto
Bridge
Balance
button;
the
fast
voltage
steps
seen
at
the
start
and
finish
of
the
current
step
should
be
eliminated.
You
may
need
to
manually
adjust
the
Bridge
Balance
M
Ω
value
for
optimum
balance.
The
M
Ω
value
is
the
resistance
of
the
electrode.
Bridge
Balance
in
the
Cell
The
Bridge
Balance
should
be
frequently
checked
when
inside
a
cell,
because
the
electrode
resistance
can
drift.
While
setting
Bridge
Balance,
Pipette
Capacitance
Neutralization
should
also
be
set
(see
Capacitance
Neutralization
).
Both
settings
can
be
monitored
continuously
through
the
experiment
by
injecting
a
small
current
step
near
the
beginning
of
each
data
sweep.
It
is
recommended
that
Pipette
Capacitance
Neutralization
be
set
at
the
same
time
as
Bridge
Balance,
because
both
the
electrode
capacitance
and
the
electrode
resistance
cause
errors
if
left
uncompensated.
Also,
it
is
easier
to
correctly
balance
the
bridge
when
electrode
capacitance
is
minimized,
because
the
“break”
between
the
rapidly
decaying
voltage
across
the
microelectrode
and
the
slowly
decaying
voltage
across
the
cell’s
membrane
resistance
is
more
distinct.
The
balancing
procedure
is
the
same
as
in
the
bath,
except
that
the
trace
appears
more
rounded
because
of
the
time
constant
of
the
cell
membrane.
Because
the
Tuning
pulse
width
is
typically
brief
compared
with
the
membrane
time
constant,
the
voltage
response
looks
like
a
series
of
ramping
straight
lines.
To
make
it
easier
to
see
the
fast
voltage
step
in
V
p
on
an
oscilloscope,
it
is
recommended
that
the
scope
input
be
AC
coupled
to
remove
the
resting
membrane
potential
from
the
signal.
The
scope
gain
can
118