Daily Operations
2-54
2.6 Table Pad
Each time before using the pad, check it carefully. Do not use the worn or
damaged pad and replace damaged pad immediately.
Liquid can penetrate damaged pad and cause hygienic problems.
When disinfecting the patient’s skin, make sure the disinfectant does not
accumulate under the patient.
2.6.1 Foam Pad
2.6.1.1 Installing Pad
1. Lay the pad on the table.
2. Make sure that the Velcro strap (loop) on the pad is fully aligned with that on the table top
(hook).
3. Check whether the pad is reliably installed.
Figure 2-44 Installing pad
Velcro strap
Содержание HyBase 8300
Страница 1: ...Operator s Manual HyBase 8300 HyBase 8500 Operating Table ...
Страница 2: ......
Страница 84: ...This page is intentionally blank ...
Страница 106: ...This page is intentionally blank ...
Страница 110: ...This page is intentionally blank ...
Страница 122: ...This page is intentionally blank ...
Страница 132: ...This page is intentionally blank ...
Страница 142: ...This page is intentionally blank ...
Страница 150: ...Appendices 10 8 3 Back plate up down up 90 5 down 45 5 ...
Страница 153: ...Appendices 10 11 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 158: ...Appendices 10 16 3 Back plate up down up 90 5 down 45 5 ...
Страница 161: ...Appendices 10 19 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 167: ...Appendices 10 25 If you want to know more information please see the accessories brochure for details ...
Страница 175: ...This page is intentionally blank ...
Страница 176: ...P N 046 006753 00 2 ...